Grech V*, Felice H**, Fenech A**,
DeGiovanni JV***. Amplatzer ASO device closure of secundum atrial
septal defects and patent foramen ovale. Images Paediatr
Cardiol 2003;15:42-66
|
|
*
|
Editor-in-Chief, Images Paediatr Cardiol |
|
**
|
Consultant Cardiologist, St. Luke’s Hospital, Malta |
|
***
|
Consultant Paediatric Cardiologist, Birmingham Chilren’s Hospital, UK |
| Heart Defects, Congenital | Embolization, Therapeutic, Instrumentation | Heart Catheterization/Instrumentation |
| Heart Septal Defects | Atrial/Pathology/Therapy |
Abstract
Transcatheter closure of atrial septal defects and patent foramen ovale
is now a well established therapeutic option. In this paper, we illustrate,
step by step, the technique of Amplatzer ASO closure of these defects.
Article
Introduction
Atrial septal defects
Atrial septal defects (ASDs) are congenital deficiencies in the interatrial
septum, and are most commonly found in the fossa ovalis. These lesions
comprise 10% of all congenital heart defects.
The haemodynamic effects of ASDs are related to the degree of left to right shunting and its duration. Large defects cause volume overload of the right side of the heart with pulmonary hypertension, that is initially reversible, and later irreversible, with shunt reversal (right to left). At this stage, usually in middle age, survival is limited, and such individuals become cyanosed with reduced exercise tolerance and are at risk of atrial arrhythmias that may precipitate frank heart failure.
Small ASDs often close spontaneously in the first few years of life. Larger defects seldom do. Significant ASDs (Qp/Qs of > 1.5) are therefore closed surgically or by interventional catheter techniques before school age or at the time of diagnosis later in life.
Patent foramen ovale
In utero, the foramen ovale serves as a physiologic conduit for right
to left shunting. After birth, the left atrial pressure is higher than
that in the right atrium, resulting in functional closure of the foramen
ovale, and later in anatomical closure. A patent foramen ovale (PFO) is
a persistent communication due to lack of anatomical closure.
A patent foramen ovale (PFO) is found in almost 50% of cryptogenic strokes and transient ischaemic attacks (TIA). In these individuals, the recurrence risk for stroke is 2-4% per annum, and for TIA the risk is 1-2% per annum. There is no consensus with regard to the ideal type or length of medical treatment for the prevention of such events. Indeed, it has been shown that there are no differences in risk reduction between different modes of anticoagulant or antiplatelet therapy with respect to recurrent stroke or TIA. On the other hand, a recent study with a mean of 15 months of followup showed that device closure of a PFO in individuals who have sustained a neurological event resulted in no strokes and a 1.7% per annum risk of TIA.1
A wide variety of devices for transcatheter closure of ASDs and PFOs have been developed in the past few decades. We illustrate the technique of ASD/PFO closure using the Amplatzer ASO (atrial septal occluder) device (AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427, USA).
This particular device consists of a collapsible meshwork of NITINOL (Nickel Titanium National Ordinance Ltd), an alloy that returns a device to its original shape after being collapsed into a sheath for delivery within the heart. The device shape is a self-expandable, double disk and the two disks are linked together by a connecting waist. Both discs and waist are filled with a polyester fabric in order to ensure complete closure of the defect.
Method
Transoesophageal echocardiography assessment of the defect is
crucial
| The catheter is passed through the defect into the left atrium (LA) and is wedged in the left upper pulmonary vein (LUPV), pointing upward or downward. If it proves difficult to cross to the LA (e.g. very small PFO), use an exchange wire over the MPA catheter to probe the interatrial septum with the tip of the wire. Alternatively, a transeptal catheter may be used to cross to the LA with an exchange wire. An exchange wire is passed through the MPA catheter and wedged in a pulmonary vein for stability, preferably the left upper pulmonary vein. |
![]() |
| The catheter can be seen on TOE crossing over from RA into LA across the ASD. The MPA catheter is now removed over the wire. |
![]() |
| A suitably chosen Amplatzer sizing balloon is opened and flushed (24 / 34 mm diameter sizes – table 1). The balloon is held vertically pointing downward and is inflated with saline in order to remove all trapped air. |
![]() |

| The balloon is then completely deflated. |
![]() |
| The standard short sheath is removed over the exchange wire and the sizing balloon is inserted over the wire. |
![]() |
| The hole in the groin must be generous as the balloon catheter is 9F, and is inserted through the skin without a sheath. |
![]() |
| The defect is situated approximately at the right border of the vertebral column in straight AP view. |
![]() |
The balloon is inflated across the defect with a 25% contrast solution.
A good waist should be obtained without under/overdoing it.
![]() |
![]() |
| The waist should be acquired and measured in the posteroanterior view and in the left anterior oblique view at an angle of about 15°. A suitable image is frozen on the slave monitor. |
![]() |
| For calibration purposes, the sizing rings on the balloon shaft are spaced 10mm, 5mm and 2mm apart. |
![]() |
| TOE is used to check that the defect is completely closed with the balloon, that no other defects in the interatrial septum (IAS) have been exposed, and the balloon diameter is measured. The best angle to view and measure the inflated balloon is at about 45 to 55°. |
![]() |
| The volume of fluid injected in the balloon is noted and the balloon
is now deflated. The balloon catheter is removed leaving the wire wedged
in the LUPV. The balloon is now reinflated using the same volume as was
used when the balloon was inflated across the defect.
The balloon is now passed through the template sizing holes and the waist obtained is compared with that of the image on the slave monitor. Visually check that waist obtained is about the same as that seen on the slave monitor. |
![]() |
| In this way, the diameter of the defect (the waist of the inflated balloon) has been estimated by three methods: by radiology, by TOE, and externally. |
![]() |
| The appropriate sheath (table 2) and introducer pack are opened (sizes 45 by 60/80 cm), and both sheath and introducer are flushed while still mounted on the cardboard. |
![]() |


| The loader is also flushed and the introducer is passed into the sheath. The sheath and introducer are now passed into the femoral vein, incidentally stopping bleeding from around the wire. |
![]() |
| The two piece loader is connected. |
![]() |
| The loader is now flushed through the side port. |
![]() |
| The delivery wire is inserted through the loader. |
![]() |
| The end of the delivery wire is pushed out of the other end of the loader. |
![]() |
| The Amplatzer device (table 3) is opened and the size advertised on the package labeling is checked with the template. The device is now screwed clockwise onto the delivery wire. It is then untightened slightly until a click is felt. |
|

| The device is loaded under water and can be helped manually in order to avoid splay of the end of the loader. |
![]() |
| If splay occurs, press around the end of the loader in order to decrease the degree of splay. |
![]() |
| The side port of the loader is flushed with tip of loader under water to ensure removal of all air. |
|
| The introducer is removed from the sheath, while ensuring that the sheath is pushed well into the LA. |
![]() |
| The loader is brought to the sheath. |
![]() |
| The wire is used to push the device out of the loader and into the sheath. |
![]() |
| The screw-on handle at the distal end of the wire is removed. |
![]() |
| The loader is slid off the wire. |
![]() |
| The handle is reattached. |
![]() |
| The device is pushed to the end of the sheath. |
![]() |
![]() |
|
| The LA disk and the stalk are pushed out of the sheath into the LA. |
![]() |
|
|
|
| The device is pulled back gently against the septum, and this is confirmed by feeling resistance on the wire and by TOE. |
![]() |
| The right atrial (RA) disk is now deployed in the RA. If the deployed device appears perfect on fluoroscopy and on TOE, then it is very likely that both disks have been deployed in one atrium. |
![]() |
| The device is given a ‘Minnesota wiggle’ in order to ensure that it is firmly deployed within the atrial septum. This is done by gently jerking the delivery wire in the sheath. |
|
| The device is now unscrewed off the wire. |
![]() |
| Make sure to pull back the delivery wire back into the sheath immediately as the end of the wire is rigid and sharp and may produce trauma. The deployed device is acquired on cine in the posteroanterior view and with an angulation of 30° cranial with 30° LAO. |
![]() |
![]() |
|
Points to ponder
The device size refers to the central stalk. Roughly, the actual rim
is 5mm for device sizes up to 19mm i.e. the actual device diameter is the
device size plus 10mm. For 20mm devices and larger, the rim is 7mm. It
is therefore important to ensure that the IAS can accommodate the required
device size. If necessary, the size of the entire IAS can be measured on
TOE. However, TOE foreshortens the length of the IAS, and such measurements
are better done during transthoracic echocardiography.
| The LA disk (and occasionally also the RA disk) may be deployed in a distorted manner (cobra formation). This may be due to a faulty device or the device may actually be catching on an intracardiac structure. If the latter, the device should be retrieved back in the sheath and deployed fully at a different angulation in the LA. Very occasionally, the device persists in deploying in this abnormal fashion and may have to be replaced. |
![]() |
If there are two atrial septal defects, balloon size both and assess not only the size and location of the defects, but also the distance between them. This will allow a decision as to whether to close the defects by using one oversized device or use two separate devices. If two devices are used, it is logical to first close the smaller defect so that when closing the larger defect, the larger device will overlap the smaller device.
If there is a large ASD in association with significant pulmonary stenosis, it is best to first close the ASD, leaving it attached to the delivery wire, and then balloon the pulmonary valve, releasing the device afterward. There is a significant risk of acute pulmonary oedema if pulmonary valvar stenosis is dilated before closure of the defect.
This animation summarises the procedure (figure 45).

Further reading
http://www.amplatzer.com/medical_professionals/index.html
References
![]() |
|
|
|
![]() |