|
|
![]() |
|
|
Sreeram N, DeGiovanni J. Stent implantation for
coarctation facilitated by the anterograde trans-septal approach. Images
Paediatr Cardiol 2005;22:12-17
|
| coarctation of aorta | stent therapy | |
The femoral artery was percutaneously cannulated with a 6F introducer.
A 6F endhole catheter was advanced to the descending aorta where angiography
confirmed the diagnosis of coarctation. (figures 1 and 2).

As the coarctation segment could not be crossed retrogradely with various
combinations of catheters and guidewires, the femoral vein was cannulated.
Transseptal puncture was performed using a Brockenbrough needle within
an 8F Mullins (Cook, USA) long sheath, and the sheath was advanced to the
left atrium. A 6F pigtail catheter was introduced via the sheath into the
left ventricle. The left ventricular systolic pressure was 280 mm Hg, compared
with a femoral arterial pressure of 100 mm Hg (gradient 170 mm Hg). Left
ventricular angiography demonstrated patency of the coarctation (figure
3 - arrow).

Via the catheter in the left ventricle, an 0.035" guidewire (260 cm
long) was advanced into the ascending aorta, through the coarctation and
into the descending aorta. The wire was snared using a 20mm snare device
(PFM, Germany) and exteriorised via the femoral artery, establishing a
continuous arteriovenous guidewire loop (figure 4).

Over this wire, a 14F Mullins sheath was advanced from the femoral
artery, across the coarctation and into the ascending aorta. A Palmaz P4014
stent (Johnson & Johnson, USA), mounted on a 20mm diameter balloon
(Cordis, USA) was delivered to the coarctation (figure 5), and appropriately
deployed (figures 6 and 7).

The post implantation left ventricular systolic pressure was 180 mm
Hg, with a residual gradient of 10 mm Hg. There were no procedure related
complications. The patient underwent coronary artery bypass grafting for
a severe stenosis of the left anterior descending artery 3 days later.

Discussion
Stent implantation is established therapy for coarctation of the aorta
in older children and adults.1,2 When the coarctation segment
cannot be crossed, the anterograde trans-septal approach should be used,
to confirm whether the arch is truly atretic, or still patent. In the case
described here, continuity of the arch was confirmed by left ventricular
angiography. It was possible to cross the coarctation anterogradely with
a guidewire, facilitating subsequent stent placement. Atresia of the aortic
arch can be successfully recanalised. Although balloon angioplasty has
been successfully performed following recanalisation, the use of covered
stents may avoid bleeding complications and provide a better long-term
outcome.3,4
References
|
|
|
|
|
![]() |